FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 19936034 · Received August 8, 2024

Report

Report Number
2955842-2024-17951
Event Type
Malfunction
Date Received
August 8, 2024
Date of Event
September 20, 2023
Report Date
July 17, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
10886874115661
PMA / PMN Number
K173337
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE INSTRUMENT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS; HOWEVER, THE EVALUATION IS NOT YET COMPLETE. A FOLLOW-UP MDR WILL BE SUBMITTED ONCE THE PRODUCT IS EVALUATED AND/ OR IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE VESSEL SEALER EXTEND INSTRUMENT AND COMPLETED THE DEVICE EVALUATION. THE VESSEL SEALER EXTEND INSTRUMENT WAS ANALYZED AND THE COMPLAINT WAS NOT CONFIRMED OR REPLICATED BY FAILURE ANALYSIS. A VISUAL INSPECTION DID NOT REVEAL ANY DAMAGE. THE BLADE WAS FOUND INTACT. THE INSTRUMENT WAS PLACED AND DRIVEN ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION AND ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIPS OPENED AND CLOSED PROPERLY. THE INSTRUMENT CORD WAS CUT SO A CUTTING TEST COULD NOT BE PERFORMED. NO PRODUCT ISSUE WAS IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SLEEVE GASTRECTOMY SURGICAL PROCEDURE, THE VESSEL SEALER EXTEND (VSE) INSTRUMENT HAD CUTTING ISSUE AND WAS FOUND TO HAVE NO BLADE. THERE WAS NO REPORT OF FRAGMENT(S) FALLING INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2397668 ENDOWRIST VESSEL SEALER EXTEND NAY INTUITIVE SURGICAL, INC 480422-01 L83230706 0185 10886874115661

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DA VINCI INSTRUMENTS AND ACCESSORIES