COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH
Report
- Report Number
- 6000001-2011-01059
- Event Type
- Malfunction
- Date Received
- February 16, 2011
- Date of Event
- May 7, 2010
- Report Date
- October 15, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-008-2-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS HAVE BEEN REPLACED. ADDITIONAL: THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.06.00, CATEGORIZED AS UNREMEDIATED. THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION (B)(4).
(B)(4). A BAXTER FIELD SERVICE TECHNICIAN SERVICED THIS DEVICE ON-SITE, AND AS SUCH THIS DEVICE WILL NOT BE RETURNED TO THE MANUFACTURER. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
BAXTER FIELD SERVICE TECHNICIAN SERVICED A COLLEAGUE INFUSION PUMP ONSITE FOR A BATTERY ISSUE. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. DURING BAXTER'S REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT A BATTERY DEPLETED SET ALARM OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN FOR THIS DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |