FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1993576 · Received January 24, 2011

Report

Report Number
9615739-2011-00020
Event Type
Malfunction
Date Received
January 24, 2011
Date of Event
December 30, 2010
Report Date
December 30, 2010
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND TWO CASTER SUPPORTS WERE BROKEN WHICH PREVENTED THE CASTERS FROM HOLDING WHEN IN BRAKE. THE TECH REPLACED THE CASTER SUPPORTS TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE BRAKE IS NOT HOLDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1