FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 1993535 · Received February 16, 2011

Report

Report Number
3005075853-2011-00654
Event Type
Malfunction
Date Received
February 16, 2011
Date of Event
September 30, 2010
Report Date
January 20, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EMPTY. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND LOCKED OUT. THE DEVICE IS DESIGNED TO LOCK OUT WHEN ALL THE CLIPS HAVE BEEN FIRED. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TESTING COULD BE PERFORMED TO EVALUATE THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A RIGHT HEMICOLECTOMY PROCEDURE, THE DEVICE RELEASED THE CLIPS STILL OPENED INSTEAD OF APPLYING THEM CORRECTLY ON THE VESSEL. THE CLIPS THAT HAVE FALLEN INTO THE ABDOMINAL CAVITY WERE RETRIEVED AND DISCARDED. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.