FDA Adverse Event Malfunction Summary report: N

COLLEAGUE SINGLE CHANNEL CXEPUMP V1.7

MDR report key: 1993528 · Received February 16, 2011

Report

Report Number
6000001-2011-01053
Event Type
Malfunction
Date Received
February 16, 2011
Date of Event
January 26, 2011
Report Date
January 26, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: THE REPORTED CONDITION OF FAILURE CODE 812:02 WAS CONFIRMED THROUGH THE EVENT HISTORY BUT NOT DUPLICATED DUE TO A FAULTY PUMP HEAD MODULE. THE PUMP HEAD MODULE WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 8.11.00 CATEGORIZED AS UNREMEDIATED. (B)(4)

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION. (B)(4)

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN REQUESTED TO BE RETURNED FOR EVALUATION. UPON COMPLETION OF BAXTER'S INVESTIGATION A FOLLOW UP MEDWATCH WILL BE SUBMITTED. THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. (B)(4)

Description of Event or Problem · 1

THIS IS A REPORT FROM BAXTER (B)(4) OF A COLLEAGUE INFUSION PUMP IN WHICH THE DEVICE ALARMED FAILURE CODE 812:02 AND STOPPED. THE NURSE DISCONNECTED THE PUMP TO LOOK FOR A REPLACEMENT, BUT ANOTHER PUMP WAS NOT FOUND SO THE PATIENT WAS RE-CONNECTED TO THE PUMP, THE ALARM WAS CLEARED, AND THE PUMP WORKED. NO PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE SINGLE CHANNEL CXEPUMP V1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1