ENDOPATH XCEL TROCAR
Report
- Report Number
- 3005075853-2011-00648
- Event Type
- Malfunction
- Date Received
- February 16, 2011
- Date of Event
- January 24, 2011
- Report Date
- January 24, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. HOWEVER, AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE INSUFFLATIONS ISSUES. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.
(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.(B)(4)
IT WAS REPORTED THAT DURING A LAPAROSCOPY-ASSISTED DISTAL GASTRECTOMY PROCEDURE, AIR LEAKED WHEN THEY TOOK THE FORCEPS IN AND OUT. THE ABDOMINAL AIR PRESSURE WAS 6 TO 7. ALTHOUGH THE PRESSURE WAS UP TO 9 TO 10, THE EVENT CONTINUED. THE DEVICE WAS USED AS- IS TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH XCEL TROCAR | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FORCEPS |