FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 1993485 · Received February 8, 2011

Report

Report Number
3004209178-2011-80364
Event Type
Injury
Date Received
February 8, 2011
Date of Event
January 26, 2011
Report Date
January 30, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED FOR DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 409 MG/DL. THE CUSTOMER STATED THAT HE EXPERIENCED VOMITING AND NAUSEA. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, AND PROGRAMMING ON THE INSULIN PUMP WERE CORRECT. THE BASAL AND BOLUS MATCHED THE DAILY TOTALS. RAN A FIXED PRIME TEST AND THE INSULIN EXITED. PERFORMED A HIGH PRESSURE TEST AND THE DEVICE PASSED THE TEST. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization