FDA Adverse Event Other Summary report: N

NAVIGATION SOFTWARE CRANIAL UNLIMITED

MDR report key: 1993388 · Received February 8, 2011

Report

Report Number
8043933-2011-00001
Event Type
Other
Date Received
February 8, 2011
Date of Event
January 10, 2011
Report Date
January 10, 2011
Manufacturer
BRAINLAB AG
Product Code
HAW
PMA / PMN Number
K082060
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE HOSPITAL DID NOT REPORT OR STATE ANY ADVERSE EFFECT FOR THIS PT. NEVERTHELESS, A RISK TO PT HEALTH COULD NOT BE EXCLUDED FOR THIS EVENT. ACCORDING TO BRAINLAB MEASURES ARE IN PLACE FOR THE BRAINLAB DEVICE IN QUESTION IN ORDER TO MINIMIZE THE RISK AS FAR AS REASONABLY PRACTICABLE. THERE WAS NO QUALITY OR PERFORMANCE ISSUE OR MALFUNCTION OF BRAINLAB EQUIPMENT. THE BRAINLAB DEVICE WORKS ACCORDING TO THE SPECIFICATIONS. CORRECTIVE AND PREVENTATIVE ACTIONS: SINCE THERE WAS NO QUALITY OR PERFORMANCE ISSUE OR MALFUNCTION OF BRAINLAB EQUIPMENT. ACCORDING BRAINLAB MEASURES ARE IN PLACE FOR THE BRAINLAB DEVICE IN QUESTION IN ORDER TO MINIMIZE THE RISK AS FAR AS REASONABLY PRACTICABLE. BRAINLAB INTENDS TO REMIND THE USERS IN THIS HOSPITAL OF THE RELEVANT INTRAOPERATIVE ACCURACY CHECKS. THERE ARE NO FURTHER CORRECTIVE AND PREVENTIVE ACTIONS INTENDED BY BRAINLAB AT THIS POINT IN TIME. THE MANUFACTURER OF THE NON-BRAINLAB PT HEAD FIXATION SYSTEM INVOLVED IN THIS CASE HAS BEEN INFORMED BY BRAINLAB.

Description of Event or Problem · 1

AS INFORMED BY THE HOSPITAL, ONE OF THE TWO PLANNED INTRACRANIAL BIOPSIES FOR THE SAME PT, SUPPORTED BY THE BRAINLAB NAVIGATION SYSTEM, WAS UNSUCCESSFUL AT THE FIRST ATTEMPT. THE USER VERIFIED WITH AN MRI SCAN THAT THE ACTUAL APPLIED TRAJECTORY DIFFERED FROM THE PLANNED, INTENDED TRAJECTORY FOR THE UNSUCCESSFUL BIOPSY ATTEMPT. THE INITIALLY UNSUCCESSFUL BIOPSY WAS REPEATED DIRECTLY AFTER THE RE-SCAN OF THE PT AND COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NAVIGATION SOFTWARE CRANIAL UNLIMITED IMAGE GUIDED SURGERY SYSTEM, CAS HAW BRAINLAB AG 22213

Patients

Seq Age Sex Outcome Treatment
1 Other