Narratives
Additional Manufacturer Narrative · 1
A GE REP EVALUATED THE SYSTEM AND REFORMATTED THE HARD DRIVE, RELOADED SOFTWARE, AND CLEARED THE MEMORY NODES. SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED, THE SYSTEM WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |