FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 1993259
·
Received February 7, 2011
Report
- Report Number
- 1720753-2011-01064
- Event Type
- Malfunction
- Date Received
- February 7, 2011
- Date of Event
- January 25, 2011
- Report Date
- February 7, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REP EVALUATED THE SYSTEM AND RELOADED SYSTEM SOFTWARE AND CALIBRATION FILES. SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED, THE SYSTEM DISPLAYED "CORRUPTED IMAGE" ERROR MESSAGES. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|