FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1993225
·
Received February 7, 2011
Report
- Report Number
- 1720753-2011-01021
- Event Type
- Malfunction
- Date Received
- February 7, 2011
- Date of Event
- January 19, 2011
- Report Date
- February 7, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FILAMENT WAS CALIBRATED AND THE SOFTWARE WAS INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9800 SYSTEM HAD A LOW AND HIGH MILLIAMPS ERROR AND A FILAMENT REGULATOR ERROR MESSAGE DURING A CASE. THE PROCEDURE WAS COMPLETED. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |