FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 1993203 · Received January 21, 2011

Report

Report Number
1824206-2011-00411
Event Type
Malfunction
Date Received
January 21, 2011
Date of Event
December 30, 2010
Report Date
December 30, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN ISOLATED THE ISSUE TO THE CT CABLE THAT INTERFACES THE LOGIC BOARD TO THE CONNECTOR BOARD. HE REPLACED THE CT CABLE AND THE TRENDELENBURG FUNCTIONED PROPERLY.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED WOULD NOT GO INTO TRENDELENBURG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1