FDA Adverse Event
Malfunction
Summary report: N
ADVANTA BED
MDR report key: 1993203
·
Received January 21, 2011
Report
- Report Number
- 1824206-2011-00411
- Event Type
- Malfunction
- Date Received
- January 21, 2011
- Date of Event
- December 30, 2010
- Report Date
- December 30, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN ISOLATED THE ISSUE TO THE CT CABLE THAT INTERFACES THE LOGIC BOARD TO THE CONNECTOR BOARD. HE REPLACED THE CT CABLE AND THE TRENDELENBURG FUNCTIONED PROPERLY.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE BED WOULD NOT GO INTO TRENDELENBURG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |