FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1993201 · Received February 7, 2011

Report

Report Number
1720753-2011-01018
Event Type
Malfunction
Date Received
February 7, 2011
Date of Event
January 19, 2011
Report Date
February 7, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CAMERA AND 5 VOLT POWER SUPPLY WERE ADJUSTED AND THE INTERCONNECT CABLE WAS REPAIRED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9800 SYSTEM HAD DARK IMAGES AND THE TECHNIQUE TRACKED TO MAXIMUM OUTSIDE A CASE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1