PD CYCLER 110 VOLT HOMECHOICEPRO AUTOMATED
Report
- Report Number
- 1423500-2011-02042
- Event Type
- Malfunction
- Date Received
- February 16, 2011
- Date of Event
- November 25, 2010
- Report Date
- January 26, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE PRODUCT ANALYSIS LAB EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS FOUND IN THE DEVICE LOGS. THE CAUSE OF THE IIPV WAS DETERMINED TO BE: INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE IIPV. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 3. THE PATIENT'S ULTRAFILTRATION READING WAS 1448ML, INDICATING THE HOME PATIENT (HP) DRAINED 1448ML MORE THAN THEIR PROGRAMMED FILL VOLUME OF 2300ML. THIS INFORMATION MEETS IIPV CRITERIA. PRODUCT SURVEILLANCE CONTACTED THE NURSE ON (B)(6) 2011. ACCORDING TO THE NURSE THE PATIENT DID NOT REPORT ANY SYMPTOMS OF OVERFILL. ADDITIONALLY, THE PATIENT DID NOT REPORT ANY ISSUES. THE PATIENT'S THERAPY WAS MODIFIED THE DAY AFTER THIS IIPV EVENT. THE PATIENT HAS BEEN TO THE CLINIC SINCE THIS EVENT AND HAS RESUMED THERAPY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PD CYCLER 110 VOLT HOMECHOICEPRO AUTOMATED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |