FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM

MDR report key: 1993111 · Received February 16, 2011

Report

Report Number
3005075853-2011-00636
Event Type
Malfunction
Date Received
February 16, 2011
Report Date
January 28, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH FOUR RELOADS PRESENT. THE RELOADS WERE RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLES WERE NOTED TO HAVE A PROPER B-FORMATION AND THE STAPLE LINE WAS COMPLETE. THE BATCH RECORDS WERE REVIEWED AND THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THE BATCHES WERE RELEASED FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLON PROCEDURE, THE DEVICE DEPLOYED MALFORMED STAPLES. NO OTHER INFORMATION IS AVAILABLE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA G4T211

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE