FDA Adverse Event Injury Summary report: N

ADAPTA SR

MDR report key: 1992878 · Received February 16, 2011

Report

Report Number
6000094-2011-00342
Event Type
Injury
Date Received
February 16, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
NVZ
PMA / PMN Number
P980035/S043
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT NOISE ARTIFACTS WERE CAUSING OVERSENSING ON THE LEAD AND IN THE DEVICE. IT WAS FURTHER REPORTED THAT THE LEADS ELECTRODE WAS ROTATED FROM THE HEART SURFACE. THE DEVICE AND LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

ASKU

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADAPTA SR IMPLANTABLE PULSE GENERATOR NVZ MEDTRONIC S.A. ADSR01 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R