FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1992876 · Received February 5, 2011

Report

Report Number
1720753-2011-00970
Event Type
Malfunction
Date Received
February 5, 2011
Date of Event
January 18, 2011
Report Date
February 5, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE X-RAY TUBE AND PERFORMED A FILAMENT CALIBRATION. NO PT INJURY WAS REPORTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE X-RAY TUBE INTERMITTENTLY MAKES A POPPING SOUND AND THE SYSTEM WILL NOT FLUORO. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSOCPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1