FDA Adverse Event Malfunction Summary report: N

ELECSYS ANTI-CCP

MDR report key: 19927573 · Received August 7, 2024

Report

Report Number
1823260-2024-02288
Event Type
Malfunction
Date Received
August 7, 2024
Date of Event
October 8, 2020
Report Date
September 18, 2024
Manufacturer
ROCHE DIAGNOSTICS
Product Code
NHX
UDI-DI
04015630940370
PMA / PMN Number
K081338
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SERIAL NUMBER OF THE E 801 ANALYZER IS (B)(6) . THE SERIAL NUMBER OF THE SECOND E 801 ANALYZER WAS REQUESTED, BUT NOT PROVIDED. A SAMPLE FROM THE PATIENT WAS PROVIDED FOR INVESTIGATION. AN INTERFERING FACTOR AGAINST THE STREPTAVIDIN COMPONENT OF THE ELECSYS ANTI-CCP ASSAY WAS RULED OUT. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

ON (B)(6) 2022, THE PATIENT HAD AN RF RESULT OF 11.3 IU/ML (NEGATIVE). ON (B)(6) 2023, THE PATIENT HAD AN RF IGG RESULT OF 1 U/ML (NEGATIVE), AN RF IGA RESULT OF 1 U/ML (NEGATIVE), AND AN RF IGM RESULT OF 4 U/ML (NEGATIVE). INVESTIGATIONS OF THE PATIENT SAMPLE WERE ABLE TO RULE OUT OTHER HETEROPHILIC ANTIBODY INTERFERENCE. FURTHER CLARIFICATION OF THE OBSERVED DISCREPANCIES IS NOT POSSIBLE WITH AVAILABLE METHODS AND THE CURRENT STATE OF THE ART. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT ISSUE.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THEY RECEIVED QUESTIONABLE RESULTS FOR 8 SAMPLES FROM THE SAME PATIENT TESTED WITH ELECSYS ANTI-CCP ON A COBAS E 801 ANALYTICAL UNIT. THE EIGHTH SAMPLE WAS ALSO TESTED ON A SECOND E 801 UNIT. THE ELECSYS ANTI-CCP RESULTS WERE ALL HIGH (REFERENCE RANGE = < 17 U/ML), BUT THE PATIENT HAD NO CLINICAL SYMPTOMS. THE FIRST SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 108 U/ML. THE SECOND SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 193 U/ML ON (B)(6) 2022. THE THIRD SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 216 U/ML ON (B)(6) 2022. THE FOURTH SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 185 U/ML ON (B)(6) 2023. THE FIFTH SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 184 U/ML ON (B)(6) 2023. THE SIXTH SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 172 U/ML ON (B)(6) 2023. THE SEVENTH SAMPLE RESULTED IN AN ANTI-CCP VALUE OF 184 U/ML ON (B)(6) 2023. THE EIGHTH SAMPLE INITIALLY RESULTED IN AN ANTI-CCP VALUE OF 162 U/ML ON (B)(6) 2024. THIS SAMPLE WAS REPEATED ON THE SAME E 801 SYSTEM, RESULTING IN A VALUE OF 184 U/ML. IT WAS ALSO REPEATED ON A SECOND E 801 ANALYZER, RESULTING IN A VALUE OF 187 U/ML. THE PATIENT WAS TESTED AT THREE OTHER HOSPITALS USING THE EUROIMMUN ANTI-CCP METHOD AND TWO UNKNOWN METHODS. ANTI-CCP RESULTS FOR ALL OTHER METHODS WERE NEGATIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234446 ELECSYS ANTI-CCP ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) NHX ROCHE DIAGNOSTICS ASKU 04015630940370

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown