PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT
Report
- Report Number
- 3005075853-2011-00620
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- January 26, 2011
- Report Date
- January 27, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RETURNED FULLY FIRED. IN ADDITION, 16 B-FORMED STAPLES WERE RECEIVED ON THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.
IT WAS REPORTED THAT DURING A PNEUMONECTOMY PROCEDURE, THE DEVICE WOULD NOT STAPLE BUT CUT. THE DEVICE CUT BUT THE STAPLES PLACED ONTO THE STAPLE LINE WERE NOT CLOSED. THE DEFECTIVES STAPLES WERE REMOVED. THE SURGEON PERFORMED A MANUAL SUTURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOAD-TCR75 |