FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT

MDR report key: 1992741 · Received February 15, 2011

Report

Report Number
3005075853-2011-00620
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
January 26, 2011
Report Date
January 27, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RETURNED FULLY FIRED. IN ADDITION, 16 B-FORMED STAPLES WERE RECEIVED ON THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PNEUMONECTOMY PROCEDURE, THE DEVICE WOULD NOT STAPLE BUT CUT. THE DEVICE CUT BUT THE STAPLES PLACED ONTO THE STAPLE LINE WERE NOT CLOSED. THE DEFECTIVES STAPLES WERE REMOVED. THE SURGEON PERFORMED A MANUAL SUTURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 RELOAD-TCR75