FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 1992706 · Received February 15, 2011

Report

Report Number
6000144-2011-01035
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S062
Removal / Correction Number
Z-0128-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154AWG ASKU

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization| R (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE TACHY LEAD