PENTAX
Report
- Report Number
- 9610877-2024-55945
- Event Type
- Malfunction
- Date Received
- August 7, 2024
- Date of Event
- July 17, 2024
- Report Date
- August 7, 2024
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL-1570STK IS AVAILABLE IN THE USA WITH A 510K NUMBER K162151. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVE PCB FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CCD DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE LG CONNECTOR FLUID DAMAGE, THE U/D KNOB BROKEN, THE ELECTRICAL PIN CONNECTOR CORRODED, THE INSERTION FLEXIBLE TUBE CRUSHED, AND THE REMOTE CONTROL BUTTONS CUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 552161 | PENTAX | VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE | EOB | HOYA CORPORATION PENTAX TOKYO OFFICE | VNL-1070STK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |