FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19926712 · Received August 7, 2024

Report

Report Number
9610877-2024-55945
Event Type
Malfunction
Date Received
August 7, 2024
Date of Event
July 17, 2024
Report Date
August 7, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL-1570STK IS AVAILABLE IN THE USA WITH A 510K NUMBER K162151. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVE PCB FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CCD DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE LG CONNECTOR FLUID DAMAGE, THE U/D KNOB BROKEN, THE ELECTRICAL PIN CONNECTOR CORRODED, THE INSERTION FLEXIBLE TUBE CRUSHED, AND THE REMOTE CONTROL BUTTONS CUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
552161 PENTAX VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL-1070STK

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown