FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 1992553 · Received February 15, 2011

Report

Report Number
2649622-2011-02914
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS SHOCKED, AND THE DEVICE WAS FOUND TO BE OVERSENSING T-WAVES. THE SENSITIVITY WAS MODIFIED. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935 ASKU

Patients

Seq Age Sex Outcome Treatment
1 47 YR Hospitalization| L| R 1688T COMPETITOR IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB