COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2011-01025
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- July 24, 2010
- Report Date
- January 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE CONDITION OF FAILURE CODE 802:02 WAS CONFIRMED THROUGH THE EVENT HISTORY DUE TO A FAULTY PUMP HEAD MODULE. THE PUMP HEAD MODULE WAS REPLACED. THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE CONDITION OF FC 808:02. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4)
(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA.
DURING REVIEW OF THE EVENT HISTORY IT WAS DETERMINED THAT A FAILURE CODE 808:02 OCCURRED DURING DELIVERY CAUSING AN INTERRUPTION OF DELIVERY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THIS DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.03.00 CATEGORIZED AS UNREMEDIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |