FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 1992435
·
Received February 15, 2011
Report
- Report Number
- 2649622-2011-02842
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- December 29, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND, HOWEVER BLOOD WAS NOTED ON HELIX MECHANISM; FULL LEAD RETURNED AND ANALYZED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT ATTEMPT THE LEAD WAS UNABLE TO BE SUCCESSFULLY PLACED TO GET GOOD ELECTRICAL MEASUREMENTS. IT WAS CAPTURING INTERMITTENTLY OR NOT AT ALL, AND HAD HIGH AND VARYING IMPEDANCES. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |