FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1992430 · Received February 15, 2011

Report

Report Number
2649622-2011-02839
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 31, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) DISTAL CONDUCTOR DISTORTED. ADDITIONAL FINDINGS OF DISTAL CONDUCTOR BLOOD/BODY FLUID (NOT OBSTRUCTED), BLOOD IN/ON HELIX MECHANISM (SLEEVE HEAD), BLOOD IN/ON HELIX MECHANISM, TIP SEAL OBSERVATION AND DAMAGED AT IMPLANT. ANALYST COMMENTED THAT HELIX CAN NOT EXTEND AND RETRACT DUE TO DISTAL CONDUCTOR DISTORTION WITHIN THE CONNECTOR. FULL LEAD RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN HAD DIFFICULTIES EXTENDING THE HELIX OF THE LEAD DURING THE IMPLANT PROCEDURE. THE LEAD WAS NOT USED AND A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other