FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1992396 · Received February 15, 2011

Report

Report Number
2649622-2011-02819
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 27, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): NO ANOMALIES WERE FOUND, HOWEVER BLOOD WAS NOTED ON THE HELIX; THE FULL LEAD WAS RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE ATRIAL LEAD IMPLANT PROCEDURE, AFTER EXTENDING AND RETRACTING THE LEAD HELIX SEVERAL TIMES, IT WAS NOTED THAT THE HELIX WAS "GUMMED UP", AND WOULD NOT EXTEND AND RETRACT. THE LEAD WAS REMOVED AND ANOTHER WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 72 YR Other