FDA Adverse Event Injury Summary report: N

FETAL GUARD

MDR report key: 199239 · Received November 16, 1998

Report

Report Number
199239
Event Type
Injury
Date Received
November 16, 1998
Date of Event
November 3, 1998
Report Date
November 16, 1998
Manufacturer
AIR SHIELDS - VICKERS
Product Code
HEL
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT PRESENTED IN LABOR AND WAS ASSESSED IN TRIAGE USING FETAL MONITOR. FETAL MONITOR STRIP SHOWED FETAL HEART TONES 150-160. PT ADMITTED AND MOVED TO ANOTHER ROOM WITH ANOTHER FETAL MONITOR. THIS MONITOR ALSO SHOWED FETAL HEART TONES AT 150-160 WITH OCCASIONAL BREAKS WHERE RATE DROPPED TO 70-80'S WHICH WAS ASSESSED AS MATERNAL HEART RATE. RATE ALWAYS RETURNED TO 150'S. PHYSICIAN CAME IN TO DELIVER BABY. HE RUPTURED BAG OF WATER AND APPLIED FETAL SCALP ELECTRODE - NO HEART RATE WAS ATTAINABLE BY MONITOR OR DOPTONE. AN EMERGENCY C-SECTION WAS DONE WITH DELIVERY OF STILLBORN MALE. AUTOPSY REPORTED BABY HAD BEEN DEAD IN UTERO FOR UP TO 48 HOURS PRIOR TO DELIVERY. FETAL MONITORS APPARENTLY WERE DOUBLING MATERNAL RATE FOR FETAL HEART TONE. MONITORS UNDER WARRANTY AND REMOVED FROM SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FETAL GUARD FETAL MONITOR HEL AIR SHIELDS - VICKERS FTLGRD 3000 *
2 FETALGUARD FETAL MONITOR HEL AIR SHIELDS - VICKERS FTLGRD 3000 *

Patients

Seq Age Sex Outcome Treatment
1 30 YR Hospitalization IV THERAPY PUMP