FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1992370 · Received February 15, 2011

Report

Report Number
2649622-2011-02798
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 20, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE LEAD WAS RETURNED WITH A FULLY EXTENDED HELIX AND THE HELIX EXTENDS AND RETRACTS WITHIN PRODUCT SPECIFICATION. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX DISENGAGED FROM THE HELICAL CHANNEL. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE TIP SEAL WAS OBSERVED TO BE OUT OF POSITION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE HELIX OF BOTH THE RIGHT VENTRICULAR AND THE RIGHT ATRIAL LEADS WOULD NOT EXTEND. BOTH LEADS WERE REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other