FDA Adverse Event Injury Summary report: N

KAPPA 700 DR

MDR report key: 1992363 · Received February 15, 2011

Report

Report Number
6000094-2011-00312
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035
Removal / Correction Number
Z-1510/1519-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED TO THE LAB WITH NO OUTPUT AND NO TELEMETRY. DEVICE ANALYSIS FOUND THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED DURING FOLLOW-UP. THE PATIENT HAD AN INTRINSIC RHYTHM ON THE ELECTROCARDIOGRAM WHEN A MAGNET WAS APPLIED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

ASKU

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KAPPA 700 DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. KDR701 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| O| R