FDA Adverse Event Malfunction Summary report: N

CAPSURE Z NOVUS

MDR report key: 1992307 · Received February 15, 2011

Report

Report Number
2649622-2011-02755
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P850089/S38
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THERE WAS NO CRIMP ON THE PROXIMAL CONDUCTOR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DURING THE IMPLANT, THE IMPEDANCE MEASURED WAS HIGH. THE LEAD WAS NOT USED AND A NEW LEAD WAS USED WITHOUT PROBLEMS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE Z NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5054 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other