FDA Adverse Event Malfunction Summary report: N

UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN

MDR report key: 1992292 · Received February 15, 2011

Report

Report Number
1423500-2011-02018
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
January 31, 2011
Report Date
January 31, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ONE USED SET WAS RECEIVED WITH NO CAP ON SPIKE CONNECTOR AND NO CAP ON PATIENT CONNECTOR IN REFERENCE TO CONNECTION ISSUE. FUNCTIONALLY HAND TIGHTENED A (B)(4) LAB TITANIUM ADAPTER WITHOUT DIFFICULTY. THE SET WAS TESTED UNDERWATER WITH NO LEAK NOTED. THE COMPLAINT COULD NOT BE CONFIRMED IN THE LAB. A BATCH REVIEW WAS NOT PERFORMED AS LOT INFORMATION WAS NOT KNOWN. BASED ON THE INFORMATION GATHERED DURING BAXTER'S INVESTIGATION, THE ROOT CAUSE OF THIS REPORT WAS NOT DETERMINED. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THIS INTERNATIONAL PRODUCT IS DISTRIBUTED OUTSIDE THE U.S. AND DOES NOT HAVE A 510K NUMBER, BUT IT IS BEING REPORTED AS IT IS THE SAME OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A NURSE REPORTED TO BAXTER (B)(4) THAT WHEN THE PATIENT WAS CONNECTING A TRANSFER SET TO A TWIN BAG USING THE CLEAN FLASH THERE WAS A CONNECTION ISSUE. WHEN THE LID WAS LIFTED THE SPIKE WAS NOT CONNECTED TO ANYTHING. THE PATIENT WAS INVOLVED, BUT THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1