FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 19921518 · Received August 6, 2024

Report

Report Number
2955842-2024-17510
Event Type
Malfunction
Date Received
August 6, 2024
Date of Event
July 9, 2024
Report Date
July 9, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (ERBE). SYSTEM TESTED AND VERIFIED AS READY FOR USE. INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE ERBE FOR EVALUATION. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/OR IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE INTEGRATED ELECTROSURGICAL UNIT (IESU) WAS RETURNED FOR FAILURE ANALYSIS AND THE REPORTED FAILURE WAS CONFIRMED AND REPRODUCED. THE IESU HAD NO SIGNIFICANT SCRATCHES OR DENTS AND THE FRONT BEZEL WAS NOT CRACKED. THE IESU WAS IN GOOD CONDITION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SYSTEM REFLECTED AN ERROR C-34 MESSAGE ON THE INTEGRATED ELECTROSURGICAL UNIT (IESU) OR ERBE. THE TECHNICAL SUPPORT ENGINEER (TSE) WENT THROUGH SEVERAL TROUBLESHOOTING STEPS; BUT THE ISSUE PERSISTED. THE PROCEDURE CONTINUED AS PLANNED. THERE WAS NO REPORT OF PATIENT INJURY.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1647897 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-40 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DA VINCI INSTRUMENTS AND ACCESSORIES