DAVINCI XI
Report
- Report Number
- 2955842-2024-17510
- Event Type
- Malfunction
- Date Received
- August 6, 2024
- Date of Event
- July 9, 2024
- Report Date
- July 9, 2024
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110898
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (ERBE). SYSTEM TESTED AND VERIFIED AS READY FOR USE. INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE ERBE FOR EVALUATION. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/OR IF ADDITIONAL INFORMATION IS RECEIVED.
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE INTEGRATED ELECTROSURGICAL UNIT (IESU) WAS RETURNED FOR FAILURE ANALYSIS AND THE REPORTED FAILURE WAS CONFIRMED AND REPRODUCED. THE IESU HAD NO SIGNIFICANT SCRATCHES OR DENTS AND THE FRONT BEZEL WAS NOT CRACKED. THE IESU WAS IN GOOD CONDITION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SYSTEM REFLECTED AN ERROR C-34 MESSAGE ON THE INTEGRATED ELECTROSURGICAL UNIT (IESU) OR ERBE. THE TECHNICAL SUPPORT ENGINEER (TSE) WENT THROUGH SEVERAL TROUBLESHOOTING STEPS; BUT THE ISSUE PERSISTED. THE PROCEDURE CONTINUED AS PLANNED. THERE WAS NO REPORT OF PATIENT INJURY.
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1647897 | DAVINCI XI | VISION SIDE SYSTEM | NAY | INTUITIVE SURGICAL, INC | 381121-40 | N/A | 00886874110898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose | DA VINCI INSTRUMENTS AND ACCESSORIES |