FDA Adverse Event Injury Summary report: N

SIGMA 300 S

MDR report key: 1992125 · Received February 15, 2011

Report

Report Number
6000094-2011-00302
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) : THE DEVICE WAS RETURNED, AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD HIGH OUTPUT, READ THRESHOLDS, AND POSSIBLE PREMATURE BATTERY DEPLETION. THE PACEMAKER WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 300 S IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SS303 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R