FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1992050 · Received February 15, 2011

Report

Report Number
2649622-2011-02692
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND THE DEFIBRILLATION COIL WAS DISTORTED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY, THE OUTER TUBING OVERLAY MELTED, THE OUTER TUBING OVERLAY HAD A COSMETIC ENVIRONMENTAL STRESS CRACK, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THE TIP SEAL WAS OBSERVED TO BE OUT OF POSITION AND THERE WAS APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD HIGH/UNSTABLE/UNMEASURABLE THRESHOLDS, HAD NO CAPTURE, WAS OVERSENSING AND DELIVERED INAPPROPRIATE SHOCKS. THE LEAD WAS REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| L| R 5076 IMPLANTABLE PACING LEAD| 6935 IMPLANTABLE TACHY LEAD| 4194 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB