FDA Adverse Event Injury Summary report: N

SIGMA 200 SR

MDR report key: 1991961 · Received February 15, 2011

Report

Report Number
6000094-2011-00298
Event Type
Injury
Date Received
February 15, 2011
Date of Event
December 19, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD SYNCOPE AND IT WAS NOT POSSIBLE TO INTERROGATE THE PACEMAKER. THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 200 SR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SSR203 ASKU

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| R