FDA Adverse Event Malfunction Summary report: N

PROXIMAL HUMERUS, RIGHT, LONG, ÿ 10X240MM

MDR report key: 19919520 · Received August 6, 2024

Report

Report Number
0009613350-2024-00329
Event Type
Malfunction
Date Received
August 6, 2024
Date of Event
July 12, 2024
Report Date
August 16, 2024
Manufacturer
ZIMMER GMBH
Product Code
HSB
UDI-DI
00889024505971
PMA / PMN Number
K200814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, G3, G6, H2, H6, H11 NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS CONFIRMED NO ABNORMALITIES OR DEVIATIONS. REVIEW OF COMPLAINT HISTORIES IDENTIFIED ADDITIONAL SIMILAR COMPLAINTS FOR THE REPORTED PRODUCTS AND NO ADDITIONAL SIMILAR COMPLAINTS FOR THE REPORTED PART AND LOT COMBINATIONS. IT CAN BE ASSUMED THAT MULTIPLE CONTRIBUTING FACTORS, RELATED TO EITHER PATIENT CONDITION AND BEHAVIOR OR IMPLANTATION PROCEDURE, CONTRIBUTED TO THE MIGRATION OF THE SCREWS. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. AN ADDITIONAL DEEPER INVESTIGATION WAS PERFORMED WHICH IDENTIFIED THE DESIGN LIMITATION OF THE CORELOCK MECHANISM OF THE AFFIXUS NAIL AS A POTENTIAL CONTRIBUTING FACTOR. AS PART OF THE DEEPER INVESTIGATION, A SCIENTIFIC LITERATURE SEARCH WAS CONDUCTED AND A SYSTEMATIC REVIEW OF THE OUTCOME OF INTRAMEDULLARY NAILING FOR ACUTE PROXIMAL HUMERUS FRACTURE SHOWED THAT SCREW MIGRATION AND PERFORATION INTO THE JOINT IS THE SECOND MOST COMMON COMPLICATION FOLLOWING SECONDARY LOSS OF REDUCTION. IN ADDITION, SCIENTIFIC LITERATURE OF COMPETITOR AND PREDICATE DEVICES WERE ASSESSED. THIS REVIEW HAS SHOWN THAT THE AFFIXUS® NATURAL NAIL® PROXIMAL HUMERAL SYSTEM PERFORMS BETTER AND/OR IN EQUAL RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

(B)(4). D10: 47248604040, ANN BLUNT TIP SCREW 4X40MM, 3193667. 47248604240, ANN BLUNT TIP SCREW 4X42MM, 3174923. 47248604440, ANN BLUNT TIP SCREW 4X44MM, 3177548. 47248612440, CORTICAL BONE SCREW, ÿ 4X24MM, 3191015. 47248612640, CORTICAL BONE SCREW, ÿ 4X26MM, 3185183. 47248612640, ANN CORT BONE SCREW 4 X 26MM, 3193252. 47248613040, CORT BONE SCREW 4 X 30MM, 3193269. 47248801002, AFFIXUS PH NL CAP 10.5X2.5MM, 3189232. G2: REPORT SOURCE JAPAN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT 4 WEEKS AFTER THE INITIAL SURGERY, SURGEON FOUND #2 PROXIMAL SCREW WAS BACKED OUT FROM THE PROPER POSITION. THE SURGEON WILL KEEP AN EYE ON THE PATIENT CONDITION AS WELL, AS NO REVISION WILL BE PLANNED SO FAR. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1546534 PROXIMAL HUMERUS, RIGHT, LONG, ÿ 10X240MM TRAUMA, INSTRUMENTS. HSB ZIMMER GMBH N/A 3179856 00889024505971

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown SEE H11 NARRATIVE.