FDA Adverse Event
Injury
Summary report: N
ATTAIN OTW
MDR report key: 1991875
·
Received February 15, 2011
Report
- Report Number
- 2649622-2011-02599
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S012
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS STRETCHED, THE INNER INSULATION WAS TORN, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD DISLODGED. THE LEAD WAS EXPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4194 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| R | 6949 IMPLANTABLE TACHY LEAD| 5568 IMPLANTABLE PACING LEAD| 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIB |