FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOG BC

MDR report key: 19918498 · Received August 6, 2024

Report

Report Number
1710034-2024-00826
Event Type
Malfunction
Date Received
August 6, 2024
Date of Event
July 9, 2024
Report Date
August 20, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825233
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. E1. ADDRESS INFORMATION WAS NOT PROVIDED, THEREFORE, XX WAS USED AS A PLACE HOLDER.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382523 AND LOT NUMBER 4030768. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOG BC CATHETER SPLIT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IV INSERTION ATTEMPTED BUT MALFUNCTIONED AND IV CATHETER SPLIT AT THE END BUT REMAINED INTACT NO IMPACT TO PATIENT/HCP.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1703387 BD INSYTE AUTOG BC PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4030768 00382903825233

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown