FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1991830 · Received February 15, 2011

Report

Report Number
2649622-2011-02571
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. 1 - PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD IMPEDANCE ON (B)(6) 2010 09:00:16. DAILY PACE IMPEDANCE TREND DATA SHOWS AN ABRUPT INCREASE FOR VENTRICULAR PACE BI IMPEDANCE=988 TO 4047 OHMS PEAK BETWEEN (B)(6) 2010 AND (B)(6) 2010. 2 - VENTRICULAR NST (NON-SUSTAINED TACHYCARDIA) =210 MS AVERAGE V-CYCLE BETWEEN (B)(6) 2010 16:53:13 AND (B)(6) 2010 10:34:08. 1 - VF=200 MS ON (B)(6) 2010 13:46:16. 1 - PATIENT ALERT FOR LEAD FAILURE PREDICTOR ON (B)(6) 2010 09:00:16. WEEKLY LOG DATA IN SAVE TO DISK FILE (B)(4) SHOWS SIX WEEKS OF MISSING IMPEDANCE MEASUREMENT DATA FOR PACE, AND DEFIB IMPEDANCE BETWEEN (B)(6) 2010 AND (B)(6) 2010.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) TRIGGERED, AND THE LEAD EXHIBITED HIGH IMPEDANCE AND OVERSENSING. THE LEAD WAS CHECKED USING THE ANALYZER, AND NO ISSUES WERE FOUND. A BACKUP PACEMAKER WAS IMPLANTED, AND THE LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization| R (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD