SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-02569
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS OVERSENSING THE T-WAVE WHILE THE PATIENT WAS EXERCISING AND THE PATIENT RECEIVED INAPPROPRIATE THERAPY. IT WAS LATER REPORTED THAT THE T-WAVE OVERSENSING WAS OBSERVED WHEN THE PATIENT WAS HAVING PREMATURE VENTRICULAR CONTRACTIONS. THE LEAD IS STILL-IN-USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS OVERSENSING THE T-WAVE WHILE THE PATIENT WAS EXERCISING AND THE PATIENT RECEIVED INAPPROPRIATE THERAPY. THE LEAD IS STILL-IN-USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Life Threatening| O | 7232CX IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 7232CX IMPLANTABLE PACEMAKER/CARDIO/DEFIB |