FDA Adverse Event Malfunction Summary report: N

VIRTUOSO II DR

MDR report key: 1991799 · Received February 15, 2011

Report

Report Number
6000144-2011-00937
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 16, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S157
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS FEELING DIZZY. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO II DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D274DRG ASKU

Patients

Seq Age Sex Outcome Treatment
1 61 YR Other (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACING LEAD