FDA Adverse Event
Malfunction
Summary report: N
ATTAIN ABILITY
MDR report key: 1991793
·
Received February 15, 2011
Report
- Report Number
- 2649622-2011-02556
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE EXISTING LEFT VENTRICULAR LEAD HAS HIGH THRESHOLDS. AN ATTEMPT TO IMPLANT TWO DIFFERENT LEFT VENTRICULAR LEADS WERE UNSUCCESSFUL DUE TO PATIENT ANATOMY. THE EXISTING LEFT VENTRICULAR LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN ABILITY | IMPLANTABLE PACING LEAD | OJX | MEDTRONIC PUERTO RICO, INC. | 4196 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Other | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6940 IMPLANTABLE PACING LEAD| 6932 IMPLANTABLE TACHY LEAD| 6937A IMPLANTABLE TACHY LEAD |