DUAL MOBILITY VIVACIT E BEARING
Report
- Report Number
- 0001822565-2024-02580
- Event Type
- Injury
- Date Received
- August 6, 2024
- Date of Event
- July 17, 2024
- Report Date
- September 5, 2024
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- LPH
- UDI-DI
- 00889024572706
- PMA / PMN Number
- K190656
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). D10: CAT# 00877502802 LOT# 3101747 BIOLOX® DELTA, CERAMIC FEMORAL HEAD THE CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, AS ITS LOCATION IS UNKNOWN. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. MEDICAL RECORDS WERE NOT PROVIDED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. THE COMPLAINT CANNOT BE CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT APPROXIMATELY TWO YEARS POST-IMPLANTATION, THE PATIENT UNDERWENT A RIGHT HIP REVISION DUE TO INSTABILITY. THE BEARING AND HEAD WERE REVISED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1562639 | DUAL MOBILITY VIVACIT E BEARING | PROSTHESIS, HIP | LPH | ZIMMER BIOMET, INC. | N/A | 65311975 | 00889024572706 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male | Hospitalization| R |