FDA Adverse Event Injury Summary report: N

MARQUIS DR

MDR report key: 1991720 · Received February 15, 2011

Report

Report Number
6000094-2011-00274
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
P980016/S23
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4) THE LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DEFIBRILLATION COIL WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY, OUTER TUBING OVERLAY MELTED, OUTER TUBING OVERLAY COSMETIC ENVIRONMENTAL STRESS CRACKING, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT REPORTED TO THE HOSPITAL AND HAD TO BE "REANIMATED". IT WAS ALSO REPORTED THAT THERE WAS HIGH IMPEDANCE, OVERSENSING, NOISE, AND AN APPARENT LEAD FRACTURE ON THE VENTRICULAR LEAD. IT WAS ALSO REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY. MYOCARDITIS AND PLEURAL EFFUSION ALSO REPORTED. THE DEVICE AND LEADS WERE EXTRACTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

ASKU

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MARQUIS DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. 7274 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R