FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 1991715 · Received February 15, 2011

Report

Report Number
2649622-2011-02516
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED) AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD IMPEDANCE WAS HIGH, ITS THRESHOLD INCREASED, AND THERE WAS A DECLINE IN THE R-WAVE AMPLITUDE. IT WAS ALSO REPORTED THAT THE LEAD FRACTURED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6944 ASKU

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB