SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2011-02499
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): DISTAL CONDUCTOR FRACTURED, FULL LEAD WAS RETURNED AND ANALYZED. IT WAS ALSO NOTED THAT THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS TORN, AND THE HELIX WAS BENT. BLOOD/BODY FLUID WAS ALSO NOTED ON THE DISTAL CONDUCTOR, THE HELIX MECHANISM SLEEVE HEAD, AND THE HELIX MECHANISM.
IT WAS REPORTED THE PATIENT RECEIVED INAPPROPRIATE SHOCKS, THE LEAD HAD INCREASING IMPEDANCE AND THERE WAS A SUSPECTED LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |