FDA Adverse Event Injury Summary report: N

ASKU

MDR report key: 1991672 · Received February 15, 2011

Report

Report Number
2183613-2011-00020
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS COULD NOT CONFIRM THE REPORTED EVENT. THE BATTERY RELEASE, BATTERY DRAWER, BATTERY DRAWER O-RING, AND BATTERY FLEX WERE CONTAMINATED, AND THE BATTERY CONTACTS WERE COMPRESSED. ONE SIDE BAIL WAS MISSING, AND THE UPPER CASE, LOWER CASE, AND TWO SIDE BAIL COVERS WERE BROKEN.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE MALFUNCTIONED, AND THE PATIENT WENT ASYSTOLIC. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION WERE UNSUCCESSFUL.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE MALFUNCTIONED, AND THE PATIENT WENT ASYSTOLIC. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| O