FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 1991667 · Received February 15, 2011

Report

Report Number
6000144-2011-00911
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Removal / Correction Number
Z-0126-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED AND THE DEVICE HAS BEEN EXPLANTED AND REPLACED. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A DECREASE IN BATTERY VOLTAGE AND PREMATURE BATTERY DEPLETION. DEVICE REMAINS IN USE AND NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE DEVICE WAS LATER EXPLANTED AND REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A DECREASE IN BATTERY VOLTAGE AND PREMATURE BATTERY DEPLETION. DEVICE REMAINS IN USE AND NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| R 4076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| 6947 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD