FDA Adverse Event Malfunction Summary report: N

ENTRUST AT

MDR report key: 1991624 · Received February 15, 2011

Report

Report Number
6000144-2011-00903
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 20, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. POWER ON RESET - POWER ON RESET PARAMETERS ONE - POR FOR CRITICAL RAM PARITY ERROR, ADDRESS=1493, DATA=4D ON (B)(6) 2010 16:05:45. ONE - PATIENT ALERT FOR DEVICE CIRCUIT ERROR ON (B)(6) 2010 16:05:45. THE DEVICE WAS NOT RETURNED, BUT DIAGNOSTIC INFORMATION IS CONSISTENT WITH THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD A POWER-ON RESET. THE DEVICE REMAINS IN USE AND NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154ATG ASKU

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD