FDA Adverse Event
Malfunction
Summary report: N
VIRTUOSO II DR
MDR report key: 1991613
·
Received February 15, 2011
Report
- Report Number
- 6000144-2011-00901
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- December 10, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S157
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THERE WAS NOISE ON THE ATRIAL CHANNEL AT IMPLANT. SENSING DIFFICULTY WAS ALSO REPORTED. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO II DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D274DRG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Other | (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD| (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD |